8972 - HIOKI DC/RMS Unit for MR8847
All products ship from Toronto, Canada by courier. Shipping is usually 1-10 days depending on your location in Canada. Shipping is free of charge. Orders received before 2:00pm (Mon-Fri) will be shipped on the same day (provided item is available in stock). You can pay with all major credit cards. On our side, we use PayPal to process all credit card transactions. But on your side, it will be like any regular purchase. You do not need to create a PayPal account. From the shopping cart, simply select "PayPal / Credit Card", then select "Pay with a credit or Visa debit card". Customer satisfaction guaranteed, or your money back. RCC Electronics is proudly Canadian. We have been selling the best test equipment since 1978. |
Instantly measure waveforms and RMS graphs from DC to AC, isolated input Key Features • 12-bit A/D resolution • From low to high frequency, wide bandwidth RMS rectifier installed(DC, or 30 Hz to 100 kHz) • Monitor instantaneous waveforms without the need to pass through the rectifying circuit • Isolated input, up to 300 V AC/DC from the earth
Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Measurement functions | Number of channels: 2, for voltage measurement, DC/RMS selectable |
---|---|
Input connectors |
Isolated BNC connector (input impedance 1 MΩ, input capacitance 30 pF), Max. rated voltage to earth: 300 V AC, DC (with input isolated from the unit, the maximum voltage that can be applied between input channel and chassis and between input channels without damage) |
Measurement range |
5 mV to 20 V/div, 12 ranges, full scale: 20 div, (AC voltage for possible measurement/display using the memory function: 280 V rms) Low-pass filter: 5/50/500 Hz, 5 k/100 kHz |
Measurement resolution | 1/100 of measurement range (using 12-bit A/D conversion and when installed in 8847) |
Highest sampling rate | 1 MS/s (simultaneous sampling across 2 channels) |
Measurement accuracy | ±0.5 % of full scale (with filter 5 Hz, zero position accuracy included) |
RMS measurement |
RMS amplitude accuracy: ±1 % of full scale (DC, 30 Hz to 1 kHz), ±3 % of full scale (1 kHz to 100 kHz), Response time: SLOW 5 s (rise time from 0 to 90% of full scale), MID 800 ms (rise time from 0 to 90% of full scale), FAST 100 ms (rise time from 0 to 90% of full scale), Crest factor: 2 |
Frequency characteristics | DC to 400 kHz -3 dB, with AC coupling: 7 Hz to 400 kHz -3dB |
Input coupling | AC/DC/GND |
Max. allowable input | 400 V DC (the maximum voltage that can be applied across input pins without damage) |
Dimensions and mass | 106 mm (4.17 in) W × 19.8 mm (0.78 in) H × 196.5 mm (7.74 in) D, 250 g (8.8 oz) |
Accessories | None |
Write a review
Your Name:Your Review: Note: HTML is not translated!
Rating: Bad Good
Enter the code in the box below: