All products ship from Toronto, Canada by courier. Shipping is usually 1-10 days depending on your location in Canada. Shipping is free of charge. Orders received before 2:00pm (Mon-Fri) will be shipped on the same day (provided item is available in stock). You can pay with all major credit cards. On our side, we use PayPal to process all credit card transactions. But on your side, it will be like any regular purchase. You do not need to create a PayPal account. From the shopping cart, simply select "PayPal / Credit Card", then select "Pay with a credit or Visa debit card". Customer satisfaction guaranteed, or your money back. RCC Electronics is proudly Canadian. We have been selling the best test equipment since 1978. |
■ Kit includes
LR8431-20 - Recorder
9782 - Case
9780 - Battery pack
9728 - Memory card (512MB)
■ Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)
Number of channels |
Analog: 10 isolated channels using scanning input method (M3 mm dia. screw terminal block) Pulse: 4 channels (All pulse inputs share common ground with the main unit) |
---|---|
Measurement parameters |
Voltage : ±100 mV to ±60 V, 1-5V f.s. 6 ranges, Max. resolution 5µV Temperature (thermocouples) : -200 ˚C to 1800 ˚C (depend on the sensor), 1 range (K, J, E, T, N, R, S, B), Max. resolution 0.1 ˚C Temperature (Pt 100 sensor) : not available Humidity : not available Totalized pulses : 0 to 1000M pulse, 1 range (No-voltage ‘a’ contact, open collector or voltage input), Max. resolution 1 pulse Rotation count : 0 to 5000/n (r/s) f.s. 1 range (No-voltage ‘a’ contact, open collector or voltage input), Resolution 1/n (r/s) Note: n = pulses per rotation (1 to 1,000) |
Max. allowable input | DC 60 V (Analog input), DC -5 V to 10 V (Pulse input) |
Max. rated voltage to earth | AC 30 Vrms, DC 60 V (Upper limit voltage that does not cause damage when applied between input channel and chassis, and between each input channels) |
Recording intervals | 10 ms to 1 hour, 19 selections (All input channels are scanned within each recording interval.) |
Selectable filters | 50 Hz, 60 Hz, or OFF (digital filtering of high frequencies on analog channels) |
Memory capacity | Internal storage: 3.5 M-words, External storage: CF card or USB memory stick (only the HIOKI CF card is guaranteed for correct operation) |
External interface | USB 2.0 mini-B receptacle ×1; Functions: Control from a PC, Transfers files from the installed CF card to a PC (cannot transfer files from the connected USB memory stick to a PC via USB communication), Data copy between CF card and USB memory stick |
Display | 4.3-inch WQVGA-TFT color LCD (480 × 272 dots) |
Functions | Save data to the CF card or USB memory stick in real time, Numerical Calculations, etc. |
Power supply |
AC adapter Z1005: 100 to 240 VAC (50/60 Hz) Battery pack 9780: Continuous use 2.5 hours 12 V DC supply: 10 to 16 V (please contact HIOKI distributor for cable; less than 3 m/9.84 ft cable length) |
Dimensions and mass | 176 mm (6.93 in) W × 101 mm (3.98 in) H × 41 mm (1.61 in) D, 550 g (19.4 oz) (Battery pack 9780 not installed) |
Accessories | Measurement Guide ×1, CD-R (Instruction manual PDF, Logger Utility Instruction Manual PDF, Data acquisition application program Logger Utility) ×1, USB cable ×1, AC adapter Z1005 ×1 |