Pass-through AC/DC current sensor rated 800 A, DC to 4 MHz, with 120 dB CMRR for high-current reactor loss and EV inverter motor R&D; built to order.
Contact RCC Electronics for Canadian availability, configuration guidance and a formal quotation.
| Rated current | 800 A AC/DC |
|---|---|
| Max. allowable input | Within the derating range, However, up to ±1200 A peak is allowable if within 20 ms. (design value) |
| Frequency bandwidth | Amplitude: DC to 4 MHz (CT6904A-2), DC to 2 MHz (CT6904A-3); Phase: DC to 1 MHz |
| Linearity | ±12.5 ppm Typical (23°C (73°F)) |
| Offset voltage | ±10 ppm Typical (23°C (73°F), no input) |
| Basic accuracy | DC (Amplitude: ±0.030 % rdg. ±0.009 % f.s., No phase specification); 45 Hz ≤ f ≤ 65 Hz (Amplitude: ±0.025 % rdg. ±0.009 % f.s., Phase: ±0.08°) |
| Output voltage rate | 2 mV / A rated (This device outputs AC+DC voltage via the Sensor Unit) |
| Max. rated voltage to earth | 1000 V CAT III |
| Core diameter | φ 32 mm (1.26 in.) |
| Operating temperature, humidity | -10°C to 50°C (14°F to 122°F); 80% RH or less (with no condensation) |
| Power supply | Power supplied from compatible Hioki power analyzers, sensor units, or current units. |
| Max. rated power | 7 VA Max. (800 A/55 Hz measurement, with a power supply of ±12 V) |
| Dimensions and mass | 139 mm (5.47 in.)W × 120 mm (4.72 in.)H × 52 mm (2.05 in.)D, CT6904A-2: 1.15 kg (40.6 oz.), cable length 3 m (9.84 ft.), CT6904A-3: 1.45 kg (51.1 oz.), cable length 10 m (32.81 ft.) |
Optional accessories below are not included unless stated on the RCCE quotation.
Confirm compatibility and the final configuration with RCC Electronics before ordering.