Brand: HIOKI
Product Code: MR8901
Availability: In Stock
Price: CAD 2,926.00
Qty:  
      

All products ship from Toronto, Canada by courier. Shipping is usually 1-10 days depending on your location in Canada. Shipping is free of charge. Orders received before 2:00pm (Mon-Fri) will be shipped on the same day (provided item is available in stock). 

You can pay with all major credit cards. On our side, we use PayPal to process all credit card transactions. But on your side, it will be like any regular purchase. You do not need to create a PayPal account. From the shopping cart, simply select "PayPal / Credit Card", then select "Pay with a credit or Visa debit card".

Customer satisfaction guaranteed, or your money back. RCC Electronics is proudly Canadian. We have been selling the best test equipment since 1978.

 

Basic specifications (Accuracy guaranteed for 1 year, Post-adjustment accuracy guaranteed for 1 year)

Functions No. of channels: 4, for voltage measurement
Input connectors Isolated BNC connector (input resistance 1 MΩ, input capacitance 10 pF)
Max. rated voltage to earth: 100 V AC, DC (with input isolated from the main unit, the max. voltage that can be applied between input channel and chassis and between input channels without damage)
Measurement range 5 mV to 10 V/div, 11 ranges, full scale: 20 div
* AC voltage can be measured/displayed: up to 140 V rms at ×1/2 amplitude compression, but limited to 100 V rms according as max. rated voltage to earth
Low-pass filter Low-pass filter: 5/50/500 Hz, 5 kHz, OFF
Resolution 1/1250 of measurement range (using 16-bit A/D converter)
Highest sampling rate 500 kS/s (simultaneous sampling across 4 channels)
Accuracy ±0.5 % of full scale (with filter 5 Hz, Zero position accuracy included)
Frequency characteristics DC to 100 kHz -3 dB
Input coupling DC/GND
Max. allowable input 150 V DC (the max. voltage that can be applied across input pins without damage)
Dimensions and mass 119.5 mm (4.70 in)W × 18.8 mm (0.74 in) H × 151.5 mm (5.96 in)D, 180 g (6.3 oz)
Accessories None

 