The HIOKI RM3545A-2 precision resistance meter measures low-resistance welds, connectors and conductors for development and production inspection. Four-terminal measurement, offset compensation and contact checking help distinguish small resistance changes.

Key Features

  • Resolution down to 1 nΩ with measurement current up to 1 A.
  • Thirteen resistance ranges from 1000 μΩ to 1000 MΩ.
  • Temperature correction with the included Z2001 sensor.
  • LAN, USB, RS-232C and EXT I/O support automated test systems.
  • Offset voltage compensation and comparator functions for production measurements.

Model configuration: RM3545A-2 accepts up to two optional Z3003 units, giving up to 20 four-wire channels. Multiplexer units are not included.

Configuration: The L2101 clip type lead is included only in the Kit version. Multiplexer units and external switch mainframes are sold separately.

See Accessories Included for the exact quantities supplied with this configuration.

Contact RCC Electronics for configuration assistance and Canadian application support.

RM3545A-2 Specifications

Instrument specifications apply under HIOKI's stated measurement conditions. Kit contents are listed separately.

Resistance range (13 ranges)[Range, Maximum display value, Resolutions, Testing current (Measurement current) ]
1000 μΩ: 1200.000 μΩ, 1 nΩ, 1 A
10 mΩ: 12.000 00 mΩ, 10 nΩ, 1 A
100 mΩ: 120.000 0 mΩ, 100 nΩ, 1 A
1000 mΩ: 1200.000 mΩ, 1 μΩ, 100 mA
10 Ω: 12.000 00 Ω, 10 μΩ, 10 mA
100 Ω: 120.000 0 Ω, 100 μΩ, 10 mA
1000 Ω: 1200.000 Ω, 1 mΩ, 1 mA
10 kΩ: 12.000 00 kΩ, 10 mΩ, 1 mA
100 kΩ: 120.000 0 kΩ, 100 mΩ, 100 μA
1000 kΩ: 1200.000 kΩ, 1 Ω, 10 μA
10 MΩ 12.000 00 MΩ, 10 Ω, 1 μA
100 MΩ (100 MΩ range high-precision mode): 120.000 0 MΩ, 100 Ω, 100 nA
1000 MΩ: 1200.0 MΩ, 100 kΩ, 1 μA or less
Representative accuracy (High mode, OVC function enabled, SLOW2, no zero adjustment)1000 μΩ range: ±0.045% rdg ±0.010 % f.s.
10 mΩ range: ±0.045% rdg ±0.001 % f.s.
100 mΩ range: ±0.045% rdg ±0.001 % f.s.
1000 mΩ range: ±0.012% rdg ±0.001 % f.s.
1000 Ω range: ±0.006% rdg ±0.001 % f.s.
Testing current (Measurement current)High mode: 1000μΩ (1 A) to 1000 MΩ (up to 1 μA)
Low mode: 100 mΩ (100 mA) to 100 Ω (1 mA)
Low power mode (LP): Low power measurement with measurement current and open circuit voltage down to 20 mV
LP1000 mΩ (1 mA) to LP1000 Ω (5 μA)
Measurement speedRepresentative value: FAST (2.3 ms) / MED (50 Hz: 22 ms, 60 Hz: 19 ms) / SLOW1 (102 ms) / SLOW2 (202 ms)
PR10 mΩ range*: FAST (21 ms) / MED (50 Hz: 41 ms, 60 Hz: 37 ms) / SLOW1 (121 ms) / SLOW2 (221 ms)
Path resistance tolerance (reference values)
Path resistance between SOURCE B and SOURCE A (other than measurement target)
Range: 100 mΩ or less (PR* mode off) : 2.6 Ω
Range: 100 mΩ or less (PR* mode on) : 3.5 Ω
Range: 1000 mΩ: 15 Ω
Range: 10 Ω : 150 Ω
Range: 100 Ω : 100 Ω
Range: 10 kΩ : 500 Ω
Range: 100 kΩ or greater: 1 kΩ
Maximum open-terminal voltageRange: 1000 Ω or less: 8.0 V
Range: 10 kΩ or greater: 20 V
Temperature measurementTemperature Sensor (Z2001[included accessories]): -10.0℃ to 99.9℃
Analog input (Ex: Infrared thermometer): DC 0 V to 2.0 V
Multiplexer (built-in option)*
*RM3545A-2 only
Multiplexer unit Z3003
Number of installable units:Max. 2
Maximum number of channels : 20 channels (4-wire method), 42 channels (2-wire method)
Switching time : 30 ms
Multiplexer (external option)Switch Mainframe
Maximum number of channels (SW1001): 33 channels (4-wire method)
Maximum number of channels (SW1002): 132 channels (4-wire method)
Switching time:11 ms
Communication interfacesLAN (TCP/IP, 10BASE-T/100BASE-TX), RS-232C (Max. 115200 bps, also used as printer interface), USB, EXT I/O (D-sub 37-pin, Analog output (D/A output voltage range: DC 0 V to 1.5 V )
FunctionsContact check, Zero adjustment (within each range ±50% f.s.) *1, Zero-adjustment-free accuracy guaranteed, OVC function, Contact improvement function (max. applied voltage: 5V; max. applied current: 10 mA), Low-power mode (maximum open voltage: 20 mV), Auto-hold function, Comparator, Temperature measurement function, Temperature correction (TC) function, Temperature conversion (ΔT) function, Statistical calculation function, Delay function, Averaging function, Saving panels (saving of measurement conditions), Data memory function, Command monitor function (display of send/receive status of commands and queries),LabVIEW® Driver compatible*2
*1: Zero adjustment forcibly disabled for 100 MΩ or greater
*2: LabVIEW Driver is the trademark or registered trademark of National Instruments.
Power supply100 V to 240 V AC, 50 Hz/60 Hz, Rated power consumption: 40 VA max.
Normal power consumption (Reference value)16 W (Testting current 1 A, LCD on)
Dimensions and massRM3545A-2: 215 mm (8.46 in) W × 80 mm (3.15 in) H × 306.5 mm (12.07 in) D, 3.4 kg (119.9 oz)

HIOKI official specifications and accessories

RM3545A-2 Kit Contents

The following quantities describe this kit configuration.

QtyModelIncluded item
1RM3545A-2precision resistance meter
1L2101Clip type lead

Standard Accessories Supplied with the Instrument

These are supplied with the main instrument. Items already counted in the kit table are not repeated.

QtyModelIncluded item
1—Power cord
1Z2001Temperature sensor
1—Male EXT I/O connector
1—EXT I/O connector cover
1—Spare fuse, F1.6AH 250 V
1—Startup guide
1—Operating precautions
1—Instruction manual

Included versus optional: The L2101 clip type lead is included only in the Kit version. Multiplexer units and external switch mainframes are sold separately.